Article

lock Open Access lock Peer-Reviewed

0

Views

ORIGINAL ARTICLE

Safety and Results of Bioelectrical Impedance Analysis in Patients with Cardiac Implantable Electronic Devices

Luíza Matos GarliniI; Fernanda Donner AlvesII; Adriano KochiI; Priccila ZuchinaliIV; Leandro ZimermanI; Mauricio PimentelIII; Ingrid Schweigert PerryIII; Gabriela Corrêa SouzaI; Nadine ClausellIII

DOI: 10.21470/1678-9741-2019-0098

ABSTRACT

Objective: To analyze the dual interference between cardiac implantable electronic devices (CIEDs) and bioelectrical impedance analysis (BIA).
Methods: Forty-three individuals admitted for CIEDs implantation were submitted to a tetrapolar BIA with an alternating current at 800 microA and 50 kHz frequency before and after the devices’ implantation. During BIA assessment, continuous telemetry was maintained between the device programmer and the CIEDs in order to look for evidence of possible electric interference in the intracavitary signal of the device.
Results: BIA in patients with CIEDs was safe and not associated with any device malfunction or electrical interference in the intracardiac electrogram of any electrode. After the implantation of the devices, there were significant reductions in BIA measurements of resistance, reactance, and measurements adjusted for height resistance and reactance, reflecting an increase (+ 1 kg; P<0.05) in results of total body water and extracellular water in liter and, consequently, increases in fat-free mass (FFM) and extracellular mass in kg. Because of changes in the hydration status and FFM values, without changes in weight, fat mass was significantly lower (-1.2 kg; P<0.05).
Conclusion: BIA assessment in patients with CIEDs was safe and not associated with any device malfunction. The differences in BIA parameters might have occurred because of modifications on the patients’ body composition, associated to their hydration status, and not to the CIEDs.

ABBREVIATIONS AND ACRONYMS

BIA = Bioelectrical impedance analysis

BCM = Body cell mass

BMI = Body mass index

CIEDs = Cardiac implantable electronic devices

CRT = Cardiac resynchronization therapy

CRT-D = Cardiac resynchronization therapy defibrillator

DBP = Diastolic blood pressure

DM = Diabetes mellitus

ECM = Extracellular mass

ECW = Extracellular water

FFM = Fat-free mass

FM = Fat mass

H = Height

HF = Heart failure

ICD = Implantable cardioverter defibrillator

ICW = Intracellular water

NYHA = New York Heart Association

PA = Phase angle

PM = Pacemaker

R = Resistance

RVI = Right ventricle impedance

RVT = Right ventricle threshold

SAH = Systemic arterial hypertension

SBP = Systolic blood pressure

SD = Standard deviation

SPSS = Statistical Package for the Social Sciences

TBW = Total body water

Xc = Reactance

INTRODUCTION

The bioelectrical impedance analysis (BIA) has been described since the 1970s as a body composition assessment method based on the opposition of the biological environment to the alternating electric current, determined by the resistance (R) and reactance (Xc) components[1,2]. It has been used both in healthy individuals and in some clinical situations, such as malnutrition, trauma, pre- and post-surgical period, liver disease, kidney failure, and heart disease[3-5]. It stands out by its applicability at bedside due to the practicality of the method, compared to other body composition techniques[6].

Cardiac implantable electronic devices (CIEDs) have been used in clinical practice for about five decades. Pacemaker (PM) is used to treat clinically significant bradyarrhythmia. Implantable cardioverter defibrillator (ICD) is used in primary and secondary prevention of sudden cardiac death. In patients with heart failure (HF), ICD and cardiac resynchronization therapy (CRT) can be used, individually or combined (CRT defibrillator [CRT-D)][7-10].

The body composition analysis through BIA assessment is not recommended for CIEDs carriers because it can cause an interference on the devices due to the applied electric current[11]. However, the small magnitude of the current sent to the body is way inferior to the susceptibility limits of the devices, such as PM and ICD. Still, there is an absence of information on the devices’ manuals about the use of BIA in CIEDs carriers.

In addition, the protocols for BIA assessment demand attention to the presence of metallic objects in order to avoid conduction of the electric current and erroneous results[11], which could be relevant to CIEDs carriers. Considering the limited amount of thorough detailed safety analyses, it is recommended to avoid applying body impedance methods in these cases[12]. This limitation may influence not only the benefits that BIA could provide for these patients, such as prognosis and body composition, but also the exclusion of this potentially critical population from analyses of body composition and mortality. This study investigated the dual interference between CIEDs and BIA.

METHODS

Study Population

This is a prospective study that included adult patients (age ≥ 18 years old) from the Brazilian Public Health System referred for CIEDs (PM, ICD, and CRT-D) implantation at the Hospital de Clínicas de Porto Alegre, from June 2015 to November 2016. Exclusion criteria were presence of hydroelectrolytic disorders, edema, amputated limbs, extreme body mass index (BMI) (<16 or >36 kg/m²), skin lesions, and women in menstrual period.

The study was approved by the Ethics Committee of the Hospital de Clínicas de Porto Alegre and it was conducted according to ethical principles of Helsinki. A written informed consent was obtained from all participants before taking part in the study.

Study Protocol

Anthropometric Evaluation

The weight values were obtained by a digital scale and the height (H) by a vertical stadiometer attached to a Filizola® scale - Personal Line (São Paulo, Brazil). Then, BMI was calculated by dividing weight (kg) by square H (m) (WHO, 2000)[13].

BIA Assessment

BIA was performed with the tetrapolar device Biodynamics 450 (Biodynamics Corp., Seattle, Washington, USA), which has a 50 kHz frequency and 800 microA current, with duration of 10 to 15 seconds. The subjects were instructed to fast for at least four hours, perform no physical activity the day prior to the assessment, and remove all metallic objects in their possession[1,14].

The assessments were performed with the patient in supine position, with arms apart from the body and separated legs. Four disposable electrocardiographic electrodes were used in each assessment (Conmed Corporation, Utica, NY). Two distal electrodes were positioned on the dorsal surfaces (in the hand, next to the metacarpus and the phalange joint, and the foot, next to the metatarsus and the phalange joint), and two were placed on the pisiform prominence of the wrist and between the medial lateral malleolus of the ankle[15]. The collected measurements of BIA were phase angle (PA), fat-free mass (FFM), fat mass (FM), body cell mass (BCM), extracellular mass (ECM), total body water (TBW), intracellular water (ICW), and extracellular water (ECW). Direct measurements of the BIA R and Xc, both also adjusted for H (R/H and Xc/H), were evaluated. BIA was performed in two moments, before and after the implantation of the CIEDs, obtaining an average of eight days (±9.2) between the first and second evaluation.

Evaluation of CIEDs Functioning

The CIEDs assessments were performed by a cardiac electrophysiologist with a specific device programmer designed for interrogation of the device’s parameters. The device programmer head was placed over the patient’s CIED to establish communication. Baseline interrogation was performed, including assessment of electrical impedance, sensing and capturing thresholds of all leads. In patients with ICD, while tachyarrhythmia detection and sensing were programmed on, all therapies (antitachycardia pacing and shock) were temporarily programmed off to avoid the possibility of inappropriate therapies. During the BIA assessment, continuous telemetry was maintained between the device programmer and CIED in order to look for evidence of possible electric interference in the intracavitary signal of the device. After completion of BIA, sensing and pacing parameters were measured again and all tachyarrhythmias therapies were programmed back to their original settings. The possible detection of anomalous electric activity secondary to the passage of the electric current not intrinsic to the heart on the ventricular electrogram was verified through continuous observation of the intracavitary signal on the device setting during the passage of the BIA’s electrical current in the presence of a high programmable sensitivity for that device.

Statistical Analysis

Data are presented in mean and standard deviation (SD) or median and interquartile range. Continuous variables were compared using Student’s t-test for paired samples or the paired Wilcoxon test, in order to obtain the differences of the BIA parameters before and after the implantation of CIEDs, and the values of the CIEDs before and after BIA. According to the BIA’s manufacturer, a 2 to 3% error is expected due to the standardization protocol for conducting the exam.

The normality analysis was assessed by the Shapiro-Wilk test. Categorical variables were described as absolute and relative frequency. The significance level used was 5%. In order to analyze the data, the Statistical Package for the Social Sciences (SPSS) software (SPSS, Chicago) version 23.0 was used.

RESULTS

The sample consisted of 43 individuals, most were male and Caucasian, mean age was 66 ± 10 years, and 39% of them presented HF (Table 1).

Table 1 - Characteristics of the study population (n=43).
Characteristics n (%); mean±SD
Gender (male) 26 (59%)
Ethnicity (Caucasian) 41 (98%)
Age (years) 66±10
SBP (mmHg) 127.5±31
DBP (mmHg) 71±15
SAH 31 (74%)
DM 13 (30%)
HF (n=17)  
    NYHA I - II 12 (71%)
    NYHA III - IV 5 (29%)
    Ischemic etiology 10 (59%)

Data described as absolute and relative frequency, mean, and standard deviation (SD). DBP=diastolic blood pressure; DM=diabetes mellitus; HF=heart failure; NYHA=New York Heart Association; SAH=systemic arterial hypertension; SBP=systolic blood pressure

Table 1 - Characteristics of the study population (n=43).

Twenty-two patients had PM implanted, 15 had an ICD, and six had CRT-D. Implanted CIEDs were manufactured by Medtronic (n=27) and Biotronik (n=16) (Table 2).

Table 2 - Characteristics of the devices implanted in the study sample.
Characteristics n (%)
Type  
    PM 22 (51%)
    Unicameral 4 (18%)
    Bicameral 18 (82%)
    ICD 15 (35%)
    Unicameral 9 (60%)
    Bicameral 6 (40%)
    CRT-D 6 (14%)
CIED manufacturer  
    Biotronik® 16 (37%)
    Medtronic® 27 (63%)
CIED implantation  
    Right side 15 (35%)
    Left side 28 (65%)

Biotronik® (Berlin, Germany); Medtronic® (Minneapolis, USA)

CIED=cardiac implantable electronic device; CRT-D=cardiac resynchronization therapy defibrillator; ICD=implantable cardioverter-defibrillator; PM=pacemaker

Table 2 - Characteristics of the devices implanted in the study sample.

Regarding variations in anthropometric and BIA measurements, weight, BMI, PA, BCM, and ICW showed no difference between the first assessment, without the implanted device, and the second, with the implanted device. However, there were significant reductions (before and after CIEDs implantations) in direct BIA measurements (R, R/H, Xc, Xc/H), reflecting an increase (+ 1 kg) in results of TBW (L) and ECW (L) and, consequently, increases in FFM (kg) and ECM (kg). Because of changes in the hydration status and FFM values, without changes in weight, FM was significantly lower (Table 3). All of these variants were approximately 1.36% of the patients' weight, being within the possible margin of error, referring to the protocol to make the examination, especially if they were changes in hydration status.

Table 3 - Comparison of the anthropometric variables and the derived variables from BIA before and after CIEDs implantation.
  Values before CIEDs implantation Values after CIEDs implantation P-value
Weight (kg) 73.4±11.0 73.3±10.8 0.941
BMI (kg/m2) 27.5±4.1 27.5±4.1 0.947
PA (º) 6.3 (5.6 - 7.0) 5.9 (5.5 - 6.9) 0.067
FFM (kg) 50.6±8.3 51.6±8.9 0.035
FM (kg) 22.8±8.3 21.6±8.1 0.017
TBW (L) 36.7±5.8 37.7±6.3 0.018
ICW (L) 19.7±4.1 19.7±4.3 0.940
ECW (L) 17.0±2.8 18.1±3.1 <0.001
BCM (kg) 23.3±4.9 23.2±4.9 0.764
ECM (kg) 27.3±4.6 28.4±4.7 0.003
R (Ohms) 483.8 (455.2 - 567.4) 464.3 (421.5 - 537.1) <0.001
R/H (Ohms/m) 317.7±64.7 299.9±59.2 0.017
Xc (Ohms) 55.3±11.9 50.7±12.2 0.001
Xc/H (Ohms/m) 33.7±7.3 31±7.3 0.001

Data described as mean and standard deviation or median and interquartile interval (P25-P75). BCM=body cell mass; BIA=bioelectrical impedance analysis; BMI=body mass index; CIEDs = cardiac implantable electronic devices; ECM=extracellular mass; ECW=extracellular water; FFM=fat-free mass; FM=fat mass; H=height; ICW=intracellular water; PA=phase angle; R=resistance; TBW=total body water; Xc=reactance

Table 3 - Comparison of the anthropometric variables and the derived variables from BIA before and after CIEDs implantation.

The impact of BIA assessment on CIEDs showed no electrical interference on intracardiac electrogram of any lead (atrial, right ventricular, left ventricular), in any patient. There was no inappropriate sensing or tachycardia detection in any device. Full interrogation of CIEDs after BIA showed no significant differences in lead parameters or device function (Table 4). No patient presented symptoms during BIA.

Table 4 - Comparison between the CIEDs parameters before and after the examination using BIA.
  Values before BIA assessment Values after BIA assessment P-value
RVI (Ohm) 547.5±97.1 550.3±107.8 0.679
R wave (mV) 11.6±7.0 12.0±6.8 0.225
RVT (V) 0.7±0.5 0.8±0.5 0.276

Data described as mean and standard deviation.

RVI=right ventricle impedance; RVT=right ventricle threshold

Table 4 - Comparison between the CIEDs parameters before and after the examination using BIA.

DISCUSSION

In our study, BIA with Biodynamics 450 (Biodynamics Corp. Seattle, Washington, USA) in patients with CIEDs was safe and not associated with any device malfunction. We did not find presence of electrical interference on intracardiac electrogram of any lead (atrial, right ventricular, left ventricular) in any patient. With respect to BIA measurements, the significant variations were related to body water and were within the margin of error of the examination.

Since this is a fast, safe, and non-invasive method, with a relatively low cost, the BIA assessment has been used to estimate the body composition and the nutritional status of healthy individuals and patients[1,16]. Nowadays, its prognostic use has been studied through PA, and a possible correlation between low PA values and mortality has been observed in patients with heart disease[4,17].

Concerning the interference of BIA on CIEDs, the same findings were observed by Buch et al.[18], while performing BIA at the frequencies of 5, 50, and 500 kHz in 20 individuals with HF. No evidence of interference with ICD function was seen in any patient, including no telemetry disruption, no oversensing on any lead, and no patient symptoms. In another larger study (n=63), which also found no electromagnetic interferences in ICDs, the authors conclude on the safety of this procedure and recommend that the current guidelines should be updated[19]. Recently, a study of the findings corroborated the results, in an analysis of 200 individuals, indicating that the use of BIA is safe in patients with cardiac devices[20].

Despite no interferences being observed on the devices in these previous studies[18,19], they did not assess the BIA parameters, which can also suffer interferences, since they are wrapped by a metallic housing. The instructions regarding the use of BIA include the removal of metallic objects, attention so that the patient is not in contact with metallic structures and, if there is presence of any sort of prosthesis, BIA examination must be on the opposite side in order to avoid conduction of the electric current[11].

In BIA measurements, an increment in TBW was observed, which may have occurred by an increase in the hydric volume due to the conditions in the hospital stay, such as saline solution administration after the procedure, and other external factors. This increase in TBW value seems to have been due to the increase in ECW (P<0.001) and for this reason there was no change in FFM (which contains large amounts of ICW ~73%). Another measure of BIA affected by hydration was the ECM, that reflects all the metabolically inactive (non-living) parts of the body, including water contained outside living cells. Still, since resistance values (R and R/H) correlate highly with fluid balance, by facilitating the passage of the electric current, the increase of TBW can impact in the reduction of R and R/H[21]. Similar results were found by Pinto et al.[22] in 62 patients with an indication for PM or defibrillator implantation, and they concluded that decrease in resistance and related parameters were associated to hydration status[2]. This limitation may influence not only the benefits that BIA could provide for these patients, such as prognosis and body composition, but also the exclusion of this potentially critical population from analyses of body composition and mortality. This study analyzed the dual interference between CIEDs and BIA[22].

A reduction in reactance values (Xc and Xc/H), reflecting tissue (not water) changes or the absorption of electrical current in tissues, was also demonstrated in our study, and this reflected in the reduction of the calculated measurement in BIA of FM.

Our study has some possible limitations that need to be considered. Our sample was relatively small, and we studied only a single BIA system in patients with CIEDs of two manufacturers. Considering these, our results cannot be applied to all patients with CIEDs or to different BIA systems. Another limitation is the variability of the evaluation time between the first and second BIA. This interval might have influenced the body composition of the individuals, as it was observed in the FM and TBW outcomes, in which case CIEDs might not have influenced this finding.

CONCLUSION

In our study, BIA assessment in patients with CIEDs was safe and not associated with any device malfunction. The differences in BIA measurements might have occurred because of modifications on the patients’ body composition, caused by their hydration status, and not by the CIEDs’ presence. Further larger studies are required to confirm these findings and to determine a change in guidelines and manufacturer’s recommendations.

REFERENCES


1. Kyle UG, Bosaeus I, De Lorenzo AD, Deurenberg P, Elia M, Gómez JM,et al. Bioelectrical impedance analysis - part I: review of principles andmethods. Clin Nutr. 2004;23(5):1226-43.doi:10.1016/j.clnu.2004.06.004.

2. Sánchez-Iglesias A, Fernández-Lucas M, Teruel JL. The electricalbasis of bioimpedance. Nefrologia. 2012;32(2):133-5.doi:10.3265/Nefrologia.pre2012.Jan.11310.

3. Mushnick R, Fein PA, Mittman N, Goel N, Chattopadhyay J, Avram MM.Relationship of bioelectrical impedance parameters to nutrition and survival inperitoneal dialysis patients. Kidney Int Suppl. 2003;64(87):S53-6.doi:10.1046/j.1523-1755.64.s87.22.x.

4. Alves FD, Souza GC, Clausell N, Biolo A. Prognostic role of phaseangle in hospitalized patients with acute decompensated heart failure. ClinNutr. 2016;35(6):1530-4. doi:10.1016/j.clnu.2016.04.007. [MedLine]

5. Llames L, Baldomero V, Iglesias ML, Rodota LP. . NutrHosp. 2013;28(2):286-95. doi:10.3305/nh.2013.28.2.6306.Spanish.

6. Mulasi U, Kuchnia AJ, Cole AJ, Earthman CP. Bioimpedance at thebedside: current applications, limitations, and opportunities. Nutr Clin Pract.2015;30(2):180-93. Erratum in: Nutr Clin Pract. 2015;30(4):589.doi:10.1177/0884533614568155.

7. Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, et al.Prophylactic implantation of a defibrillator in patients with myocardialinfarction and reduced ejection fraction. N Engl J Med. 2002;346(12):877-83.doi:10.1056/NEJMoa013474.

8. Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, et al.Amiodarone or an implantable cardioverter-defibrillator for congestive heartfailure. N Engl J Med. 2005;352(20):225-37. Erratum in: N Engl J Med.2005;352(20):2146. doi:10.1056/NEJMoa043399.

9. Martinelli Filho M, Zimerman LI, Lorga AM, Vasconcelos JTM, Rassi AJr. . Arq Bras Cardiol. 2007;89(6):e210-38.Portuguese.

10. Ponikowski P, Voors AA, Anker SD, Bueno H, Cleland JGF, Coats AJS,et al. 2016 ESC guidelines for the diagnosis and treatment of acute and chronicheart failure: the task force for the diagnosis and treatment of acute andchronic heart failure of the European society of cardiology (ESC). Developedwith the special contribution of the heart failure association (HFA) of the ESC.Eur Heart J. 2016;37(27):2129-200. Erratum in: Eur Heart J. 2016.doi:10.1093/eurheartj/ehw128. [MedLine]

11. Kyle UG, Bosaeus I, De Lorenzo AD, Deurenberg P, Elia M, ManuelGómez J, et al. Bioelectrical impedance analysis -part II: utilization inclinical practice. Clin Nutr. 2004;23(6):1430-53.doi:10.1016/j.clnu.2004.09.012.

12. Foster KR, Lukaski HC. Whole body impedance -- what does it measure?Am J Clin Nutr. 1996;64(3 Suppl):388S-96S.doi:10.1093/ajcn/64.3.388S. [MedLine]

13. World Health Organization. Obesity: preventing and managing theglobal epidemic. Report of a World Health Organization Consultation. Geneva:World Health Organization, 2000. p. 256. WHO Obesity Technical Report Series, n.284.

14. Mialich MS, Sicchieri JF, Junior AA. Analysis of body composition: acritical review of the use of bioelectrical impedance analysis. Int J Clin Nutr.2014;2(1):1-10. doi:10.12691/ijcn-2-1-1.

15. National Institutes of Health Technology Assessment ConferenceStatement. Bioelectrical impedance analysis in body composition measurement. AmJ Clin Nutr. 1996;64(3 Suppl):524S-32S.doi:10.1093/ajcn/64.3.524S. [MedLine]

16. Lee SY, Gallagher D. Assessment methods in human body composition.Curr Opin Clin Nutr Metab Care. 2008;11(5):566-72.doi:10.1097/MCO.0b013e32830b5f23.

17. Colín-Ramírez E, Castillo-Martínez L, Orea-Tejeda A, Vázquez-DuránM, Rodríguez AE, Keirns-Davis C. Bioelectrical impedance phase angle as aprognostic marker in chronic heart failure. Nutrition. 2012;28(9):901-5.doi:10.1016/j.nut.2011.11.033.

18. Buch E, Bradfield J, Larson T, Horwich T. Effect of bioimpedancebody composition analysis on function of implanted cardiac devices. Pacing ClinElectrophysiol. 2012;35(6):681-4.doi:10.1111/j.1540-8159.2012.03377.x.

19. Meyer P, Makhlouf AM, Mondouagne Engkolo LP, Trentaz F, Thibault R,Pichard C, et al. Safety of bioelectrical impedance analysis in patientsequipped with implantable cardioverter defibrillators. J Parenter Enteral Nutr.2017;41(6):981-5. doi:10.1177/0148607116633823.

20. Chabin X, Taghli-Lamallem O, Mulliez A, Bordachar P, Jean F, Futier E, et al. Bioimpedance analysis is safe in patients with implanted cardiac electronic devices. Clin Nutr. 2019;38(2):806-11. doi:10.1016/j.clnu.2018.02.029.

21. Piccoli A, Rossi B, Pillon L, Bucciante G. A new method for monitoring body fluid variation by bioimpedance analysis: the RXc graph. Kidney Int. 1994;46(2):534-9. doi:10.1038/ki.1994.305. [MedLine]

22. Pinto LW, Gandra SV, Alves MC, Gomes I, Sternick EB. Bioelectrical impedance analysis of body composition: influence of a newly implanted cardiac device. J Electr Bioimp. 2017;8(1):60-5. doi:10.5617/jeb.4760.

Author's roles & responsibilities

LMG Substantial contributions to the conception of the work, analysis, and interpretation of data for the work; drafting the work and revising it critically for important intellectual content; final approval of the version to be published

FDA Substantial contributions to the conception of the work; final approval of the version to be published

AK Substantial contributions to the conception of the work, analysis, and interpretation of data for the work; drafting the work and revising it critically for important intellectual content; final approval of the version to be published

PZ Drafting the work or revising it critically for important intellectual content; final approval of the version to be published

LZ Drafting the work or revising it critically for important intellectual content; final approval of the version to be published

MP Substantial contributions to the interpretation of data for the work; drafting the work and revising it critically for important intellectual content; final approval of the version to be published

ISP Substantial contributions to the interpretation of data for the work; drafting the work and revising it critically for important intellectual content; final approval of the version to be published

GCS Substantial contributions to the interpretation of data for the work; drafting the work and revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published

NC Drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published

Article receive on Friday, March 8, 2019

Article accepted on Wednesday, April 24, 2019

CCBY All scientific articles published at www.rbccv.org.br are licensed under a Creative Commons license

Indexes

All rights reserved 2017 / © 2024 Brazilian Society of Cardiovascular Surgery DEVELOPMENT BY